Marquette Hellige GmbH CardioSys / MicroLab V4.X Page 41
Servicing Instructions 227 436 37 Rev. B V1.3
CardioSys/MicroLab and the printer ground strip. The value must be < 0.2
! Also
measure the enclosure leakage current.
8. Connect the printer to a wall outlet and switch it on.
9. If all measurement results are within the admissible range, replace the sticker
"Important: Do not operate the device in the vicinity of the patient (1.5 m/5 ft.)" with the
sticker "CardioSys/MicroLab ..." (refer to Labelling Sheet).